Summary

FDA records medical device Class II recall Z-0261-2025, published November 6, 2024 and initiated September 16, 2024. Recalling firm: Ambu Inc.; Status: Ongoing; Reported quantity: 1824; Lot: Lot Codes: VivaSight 2 DLT Kit 35 Fr - lot no. 1000891667 to lot. no 1001021855; VivaSight 2 DLT Kit 37 Fr - lot. no. 1000887909 to lot no. 1001021861; VivaSight 2 DLT Kit 39 Fr - lot. no. 1000916871 to lot no. 1001021917; VivaSight 2 DLT Kit 41 Fr - lot. no. 1000934084 to lot. no. 1001011434. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ambu has received complaints on Ambu® VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ambu Inc.
Distributor
US Nationwide. Canada.
Lot
Lot Codes: VivaSight 2 DLT Kit 35 Fr - lot no. 1000891667 to lot. no 1001021855; VivaSight 2 DLT Kit 37 Fr - lot. no. 1000887909 to lot no. 1001021861; VivaSight 2 DLT Kit 39 Fr - lot. no. 1000916871 to lot no. 1001021917; VivaSight 2 DLT Kit 41 Fr - lot. no. 1000934084 to lot. no. 1001011434
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Ambu® VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000) recalled?

Yes. FDA published recall Z-0261-2025 on November 6, 2024. Reported hazard: Ambu has received complaints on Ambu® VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter