Back to search
FDAFDA · Food and drugs·Recall number Z-0261-2025

Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)

Published November 6, 2024 · Recall initiated: September 16, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Ambu Inc.
Distributor
US Nationwide. Canada.
Lot
Lot Codes: VivaSight 2 DLT Kit 35 Fr - lot no. 1000891667 to lot. no 1001021855; VivaSight 2 DLT Kit 37 Fr - lot. no. 1000887909 to lot no. 1001021861; VivaSight 2 DLT Kit 39 Fr - lot. no. 1000916871 to lot no. 1001021917; VivaSight 2 DLT Kit 41 Fr - lot. no. 1000934084 to lot. no. 1001011434
Risk classification
Class II
Area (FDA)
device
Country
United States