1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or mask).
Published November 6, 2024 · Recall initiated: September 27, 2024
Summary
FDA records medical device Class II recall Z-0263-2025, published November 6, 2024 and initiated September 27, 2024. Recalling firm: Westmed, Inc.; Status: Ongoing; Reported quantity: 50,050 units; Lot: Catalog Number: S6411P UDI-DI codes: Each: 00709078012488 Case: 10709078012485 Lot Numbers: 0004294660 0004289136 0004291318 0004295361 Catalog Number: 7 5132 UDI-DI codes: Each: 00709078008405 Case: 10709078008402 Lot Numbers: 0004289133 0004289134 0004289135 0004289735 0004289736 0004289737 0004290327 0004290328 0004291315 0004291316 0004291317 0004291790 0004291791 0004292272 0004292273 0004292275 0004292717 0004293127 0004294432 0004294433 0004295069 0004295358 0004295458. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Westmed, Inc.
- Distributor
- U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV O.U.S.: N/A
- Lot
- Catalog Number: S6411P UDI-DI codes: Each: 00709078012488 Case: 10709078012485 Lot Numbers: 0004294660 0004289136 0004291318 0004295361 Catalog Number: 7 5132 UDI-DI codes: Each: 00709078008405 Case: 10709078008402 Lot Numbers: 0004289133 0004289134 0004289135 0004289735 0004289736 0004289737 0004290327 0004290328 0004291315 0004291316 0004291317 0004291790 0004291791 0004292272 0004292273 0004292275 0004292717 0004293127 0004294432 0004294433 0004295069 0004295358 0004295458
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADuraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694November 6, 2024
- FDAOlympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327UNovember 6, 2024
- FDAAmbu® VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000),November 6, 2024
- FDAAccess Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, andNovember 6, 2024
- FDAProduct Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includesNovember 6, 2024
- FDANeo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible InfantJuly 8, 2020
Questions about this recall
Was 1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or mask). recalled?
Yes. FDA published recall Z-0263-2025 on November 6, 2024. Reported hazard: Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
What should I do if I have this product?
Voluntary: Firm initiated · Letter