Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening.
Published November 21, 2012 · Recall initiated: September 28, 2012
Summary
FDA records medical device Class II recall Z-0277-2013, published November 21, 2012 and initiated September 28, 2012. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 8; Lot: Part number 390.007, Lot number 6800599. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Nationwide Distribution-including the states of CA, MN, NC, and TN.
- Lot
- Part number 390.007, Lot number 6800599
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDAProgressive Medical PMI Endoscopic Dissector with Kittner Tip (5mm x 40cm), Reorder Number: PMILK40, ProductNovember 21, 2012
- FDAHX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel systemNovember 21, 2012
Questions about this recall
Was Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening. recalled?
Yes. FDA published recall Z-0277-2013 on November 21, 2012. Reported hazard: The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
What should I do if I have this product?
Voluntary: Firm initiated · Letter