epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Published November 13, 2024 · Recall initiated: September 24, 2024
Summary
FDA records medical device Class II recall Z-0286-2025, published November 13, 2024 and initiated September 24, 2024. Recalling firm: Siemens Healthcare Diagnostics Inc; Status: Ongoing; Reported quantity: 1292 units; Lot: UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- US Nationwide. Canada.
- Lot
- UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;May 6, 2026
- FDARAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.August 6, 2025
- FDASaphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowingNovember 13, 2024
- FDAAmericot 20-01S (1/4" x 1/4") or 6mm x 6mm: Neurological sponges for use in the protection of neural tissueNovember 13, 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPointJuly 3, 2024
- FDARAPIDPoint® 500 Blood Gas Analyzer with v2.4 software for professional use in a point-of-care or laboratoryMay 17, 2017
Questions about this recall
Was epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515 recalled?
Yes. FDA published recall Z-0286-2025 on November 13, 2024. Reported hazard: Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
What should I do if I have this product?
Voluntary: Firm initiated · Letter