Americot 20-01S (1/4" x 1/4") or 6mm x 6mm: Neurological sponges for use in the protection of neural tissue during surgery. Ref: 20-01S ¿ x ¿ inches or 6mm x 6mm) Component: N/A
Published November 13, 2024 · Recall initiated: September 18, 2024
Summary
FDA records medical device Class II recall Z-0287-2025, published November 13, 2024 and initiated September 18, 2024. Recalling firm: American Surgical Company, LLC; Status: Ongoing; Reported quantity: 390 units; Lot: UDI-DI: 10749915000039 Lot Number Expiration Date: 2023/03 AQ 2028/03/31; 2023/06 AP 2028/06/30; 2023/10 BO 2028/10/31; 2023/12 AY 2028/12/31; 2024/05 AE 2029/05/31; 2024/06 AQ 2029/06/30. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product defect where the X-ray detectable barium strip (aka radio-opaque marker) becomes detached from the sponge (pattie). May delay surgical process as the surgeon may require extra time to remove the markers from the surgical field.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Surgical Company, LLC
- Distributor
- CA, IA, IL, MN, MO, NY, OH, VA Foreign: Australia, Belgium, Hungry, Italy, Netherlands, Sweden, UK Italy
- Lot
- UDI-DI: 10749915000039 Lot Number Expiration Date: 2023/03 AQ 2028/03/31; 2023/06 AP 2028/06/30; 2023/10 BO 2028/10/31; 2023/12 AY 2028/12/31; 2024/05 AE 2029/05/31; 2024/06 AQ 2029/06/30
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515November 13, 2024
- FDAiotaSOFT Plus Drive Unit, REF IM-05. The iotaSOFT Insertion System is intended to aid the surgeon in placementNovember 13, 2024
- FDASaphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowingNovember 13, 2024
- FDAMedtronic DLP Aortic Root Cannula with Flow-Guard, REF 11012; for use in cardiopulmonary bypass surgeryNovember 13, 2024
- FDACartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system thatNovember 13, 2024
- FDADeRoyal C-SECTION TRACEPACK, REF 89-9338-06November 13, 2024
Questions about this recall
Was Americot 20-01S (1/4" x 1/4") or 6mm x 6mm: Neurological sponges for use in the protection of neural tissue during surgery. Ref: 20-01S ¿ x ¿ inches or 6mm x 6mm) Component: N/A recalled?
Yes. FDA published recall Z-0287-2025 on November 13, 2024. Reported hazard: Product defect where the X-ray detectable barium strip (aka radio-opaque marker) becomes detached from the sponge (pattie). May delay surgical process as the surgeon may require extra time to remove the markers from the surgical field.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail