Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

Published November 20, 2024 · Recall initiated: October 7, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Respironics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
Lot
Model Number (UDI-DI Number): DS2200X11B (606959052017); IN2200X15B (606959056497); BL2200X15B (606959058668); BR2200X18B (606959058583); CA2200X12B (606959052321); DE2200X13B (606959058651); GB2200X15B (606959058675); ND2200X15B (606959058620); ES2200X15B (606959058644); FR2200X14B (606959058613); FX2200X15B (606959061033); EU2200X15B (606959058682); IA2200X15B (606959058606); IT2200X21B (606959058637); KR2200X15B (606959058590); TR2200X15B (606959058798); EE2200X15B (606959058705). All Serial Numbers.
Risk classification
Class I
Area (FDA)
device