FDAFDA · Food and drugs·Recall number Z-0293-2025
Trilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200X15B, IT2200X21B, KR2200X15B, TR2200X15B, EE2200X15B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Published November 20, 2024 · Recall initiated: October 7, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Respironics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
- Lot
- Model Number (UDI-DI Number): DS2200X11B (606959052017); IN2200X15B (606959056497); BL2200X15B (606959058668); BR2200X18B (606959058583); CA2200X12B (606959052321); DE2200X13B (606959058651); GB2200X15B (606959058675); ND2200X15B (606959058620); ES2200X15B (606959058644); FR2200X14B (606959058613); FX2200X15B (606959061033); EU2200X15B (606959058682); IA2200X15B (606959058606); IT2200X21B (606959058637); KR2200X15B (606959058590); TR2200X15B (606959058798); EE2200X15B (606959058705). All Serial Numbers.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States