Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.
Published November 28, 2012 · Recall initiated: September 28, 2012
Summary
FDA records medical device Class II recall Z-0298-2013, published November 28, 2012 and initiated September 28, 2012. Recalling firm: Biomet 3i, LLC; Status: Terminated; Reported quantity: 76 devices; Lot: Model Number IFOA411, Lot # 2011111333, 2011111333-S1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet 3i, LLC
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of VA, SC, HI, LA, OH, FL, IL, NJ, NC, CA, CT, MI, TN, AK, NH, NY, MA, GA, PA, and KY and the countries of United Kingdom, Spain, Germany, Canada, Japan, Singapore, Taiwan, Israel, Poland, Hong Kong, Ireland, The Netherlands, Austria, and France.
- Lot
- Model Number IFOA411, Lot # 2011111333, 2011111333-S1
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABELLATEK TITANIUM ABUTMENT 4.1MM Reference Number: EDAT4October 14, 2020
- FDACERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 6MM(P) X 6MM(H) Reference Number: IEHA466March 4, 2020
- FDACERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 5MM(P) X 3MM(H) Reference Number: IEHA353March 4, 2020
- FDACERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 4.1MM(P) X 4MM(H) Reference Number: IEHA444March 4, 2020
- FDAProduct is Full OSSEOTITE 2 Tapered Certain Implant, Model Number XIFNT413, and size 4 x 13 mm. The product isNovember 28, 2012
- FDAbioMerieux BacT/ALERT Control Module, catalog number 210147, Clinical instrument The product is most commonlyNovember 28, 2012
Questions about this recall
Was Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures. recalled?
Yes. FDA published recall Z-0298-2013 on November 28, 2012. Reported hazard: Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant.
What should I do if I have this product?
Voluntary: Firm initiated · Letter