IQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
Published November 7, 2018 · Recall initiated: June 7, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- US Nationwide Distribution
- Lot
- software versions 4.7.0 or 4.7.2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States