Access 2 Immunoassay System, Catalog Number 81600N. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Published January 10, 2018 · Recall initiated: November 6, 2017
Summary
FDA records medical device Class III recall Z-0308-2018, published January 10, 2018 and initiated November 6, 2017. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 227 units; Lot: Serial/Lot Numbers: 570005-570043, 570047-570049, 570051-570061, 570066-570071, 570073-570091, 570093-570123, 570125, 570126, 570128, 570131-570153, 570156-570203, 570205-570241, 570244-570248, 570250-570252 and 512684. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.
- Lot
- Serial/Lot Numbers: 570005-570043, 570047-570049, 570051-570061, 570066-570071, 570073-570091, 570093-570123, 570125, 570126, 570128, 570131-570153, 570156-570203, 570205-570241, 570244-570248, 570250-570252 and 512684
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAEnoxaparin Sodium, Injection 120 mg/0.8 mL, pre-filled, packaged in 10-count cartons, Rx Only, Winthrop US., aJanuary 17, 2018
- FDACADD-Legacy 1, Model 6400 Ambulatory Infusion Pump, REF 21-6400-51, PC 10008552-001, Smiths Medical ASD, Inc.January 10, 2018
Questions about this recall
Was Access 2 Immunoassay System, Catalog Number 81600N. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids. recalled?
Yes. FDA published recall Z-0308-2018 on January 10, 2018. Reported hazard: Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail