Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electronic documentation of clinical data and is designed to accept, transfer, display, calculate, store and manage clinical data. System capabilities provide the user with the ability to acquire data from the medical devices and to document, annotate, display, store, print, export and retrieve patient clinical practice information. This product is intended for professional use only.
Published November 27, 2013 · Recall initiated: March 11, 2013
Summary
FDA records medical device Class II recall Z-0337-2014, published November 27, 2013 and initiated March 11, 2013. Recalling firm: GE Healthcare It; Status: Terminated; Reported quantity: 1,132 units; Lot: Centricity Perinatal and Centricity Intensive Care Versions: 6.60, 6.60.9, 6.70.0, 6.70.1, 6.70.5, 6.70.6, 6.80.0, 6.80.1, 6.90.0, 6.90.1, 6.91.1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare It
- Distributor
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Canada, Denmark, Germany, Hong Kong, Korea, South Korea, Thailand and United Kingdom.
- Lot
- Centricity Perinatal and Centricity Intensive Care Versions: 6.60, 6.60.9, 6.70.0, 6.70.1, 6.70.5, 6.70.6, 6.80.0, 6.80.1, 6.90.0, 6.90.1, 6.91.1
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an informationSeptember 7, 2016
- FDAThe Centricity Laboratory System is intended to be an information system designed to support the clinical andFebruary 5, 2014
- FDAMerge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centersJanuary 22, 2014
- FDAThe Centricity Laboratory System is intended to be an information system designed to support the clinical andJanuary 8, 2014
- FDAVascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated KnittedNovember 27, 2013
- FDAVascu Guard, Peripheral Vascular Patch with Apex Processing. For use in peripheral vascular reconstruction.November 27, 2013
Questions about this recall
Was Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electronic documentation of clinical data and is designed to accept, transfer, display, calculate, store and manage clinical data. System capabilities provide the user with the ability to acquire data from the medical devices and to document, annotate, display, store, print, export and retrieve patient clinical practice information. This product is intended for professional use only. recalled?
Yes. FDA published recall Z-0337-2014 on November 27, 2013. Reported hazard: 1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When
What should I do if I have this product?
Voluntary: Firm initiated · Letter