FDAFDA · Food and drugs·Recall number Z-0346-2018

Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

Published January 31, 2018 · Recall initiated: November 29, 2014

Summary

FDA records medical device Class I recall Z-0346-2018, published January 31, 2018 and initiated November 29, 2014. Recalling firm: Vyaire Medical; Status: Terminated; Reported quantity: 16,670 units in total; Lot: 0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Vyaire Medical
Distributor
US Nationwide Distribution
Lot
0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation. recalled?

Yes. FDA published recall Z-0346-2018 on January 31, 2018. Reported hazard: Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter