GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Published November 29, 2023 · Recall initiated: September 29, 2023
Summary
FDA records medical device Class II recall Z-0358-2024, published November 29, 2023 and initiated September 29, 2023. Recalling firm: Instrumentation Laboratory; Status: Ongoing; Reported quantity: 2448 units; Lot: Part No. 00024307507: UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C; Part No. 00024315007: UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C; Part No. 00024330007: UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J; Part No. 00024345007: UDI-DI: 08426950082545, Lot Numbers: 323761H; Part No. 00024360007: UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Instrumentation Laboratory
- Distributor
- US States: MA, MN, NY. India.
- Lot
- Part No. 00024307507: UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C; Part No. 00024315007: UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C; Part No. 00024330007: UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J; Part No. 00024345007: UDI-DI: 08426950082545, Lot Numbers: 323761H; Part No. 00024360007: UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)October 21, 2026
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnosticOctober 21, 2026
- FDAGEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010August 12, 2026
- FDAGEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011August 12, 2026
- FDAstryker Sage PrimaFit External Urine Management System for the Female AnatomyNovember 29, 2023
- FDAVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurationsNovember 29, 2023
Questions about this recall
Was GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007 recalled?
Yes. FDA published recall Z-0358-2024 on November 29, 2023. Reported hazard: Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
What should I do if I have this product?
Voluntary: Firm initiated · Letter