GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007

Published November 29, 2023 · Recall initiated: September 29, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Instrumentation Laboratory
Distributor
US States: MA, MN, NY. India.
Lot
Part No. 00024307507: UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C; Part No. 00024315007: UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C; Part No. 00024330007: UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J; Part No. 00024345007: UDI-DI: 08426950082545, Lot Numbers: 323761H; Part No. 00024360007: UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E;
Risk classification
Class II
Area (FDA)
device
Country
United States