FDAFDA · Food and drugs·Recall number Z-0374-2021
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 53 mm Number: 00-8065-650-32
Published November 11, 2020 · Recall initiated: September 24, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- US Nationwide
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?