Back to search
FDAFDA · Food and drugs·Recall number Z-0390-2019

Cordis POWERFLEX PRO PTA Dilatation Catheter - Product Usage: Is intended to dilate stenoses in iliac, femoral, ilio-femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device is also indicated for postdilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.

Published November 14, 2018 · Recall initiated: September 5, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Product may not meet the internal manufacturing shaft subassembly burst strength specification (Failure of the Burst Shaft Test).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Cordis Corporation
Distributor
Worldwide distribution - US Nationwide in the states of AL, CA, FL, GA, IL, IN, KS, KY, LA, MA, MD, MO, MS, NC, ND, NJ, OH, PA, SC, TN, TX, and WI. Countries of Japan and Netherlands.
Lot
Catalog Number Lot Number Expiry Date 4400322X 82144115 2020-04-30; 4400602S 82144141 2020-05-31; 4400515X 82144947 2020-06-30; 4400308S 82148810 2020-11-30; and 4400508S 82148811 2020-11-30.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card