FDAFDA · Food and drugs·Recall number Z-0390-2020

ORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY, UDI: (01) 18033509855948

Published November 20, 2019 · Recall initiated: September 27, 2018

Summary

FDA records medical device Class II recall Z-0390-2020, published November 20, 2019 and initiated September 27, 2018. Recalling firm: Orthofix Srl; Status: Terminated; Reported quantity: N/A; Lot: Lot # B1207286, B1207428, B1211384, B1215263 *Not marketed in US. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Orthofix Srl
Distributor
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Lot
Lot # B1207286, B1207428, B1211384, B1215263 *Not marketed in US
Risk classification
Class II
Area (FDA)
device
Country
Italy

Related recalls

Questions about this recall

Was ORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY, UDI: (01) 18033509855948 recalled?

Yes. FDA published recall Z-0390-2020 on November 20, 2019. Reported hazard: As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

What should I do if I have this product?

Voluntary: Firm initiated · Letter