Back to search
FDAFDA · Food and drugs·Recall number Z-0390-2020

ORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY, UDI: (01) 18033509855948

Published November 20, 2019 · Recall initiated: September 27, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Orthofix Srl
Distributor
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Lot
Lot # B1207286, B1207428, B1211384, B1215263 *Not marketed in US
Risk classification
Class II
Area (FDA)
device
Country
Italy