FDAFDA · Food and drugs·Recall number Z-0396-2014
Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
Published December 4, 2013 · Recall initiated: September 7, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Nationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States