Slipped Capital Femoral Epiphysis System Product Usage: Intended for fracture fixation of large bones and large bone fragments and for slipped capital femoral epiphysis, pediatric femoral neck fractures, intercondylar femur fractures and sacroiliac joint disruptions.
Published December 4, 2013 · Recall initiated: July 31, 2013
Summary
FDA records medical device Class II recall Z-0408-2014, published December 4, 2013 and initiated July 31, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 166; Lot: Parts 03.207.001 and 03.207.008 with lot numbers: Part # Lot # 03.207.001 PE00376 PE00439 PE01391 PE01329 PE01412 PE00504 03.207.008 PE00440 PE00377 PE00513 PE01475 PE01691. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a possibility for the Guide Wires of the Slipped Capital Femoral Epiphysis System to stick in the cannulation of the drill bit and the drill bits to break during surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- USA Nationwide Distribution
- Lot
- Parts 03.207.001 and 03.207.008 with lot numbers: Part # Lot # 03.207.001 PE00376 PE00439 PE01391 PE01329 PE01412 PE00504 03.207.008 PE00440 PE00377 PE00513 PE01475 PE01691
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDASynthes Titanium Cannulated Humeral Nail Intended to aid in the alignment and stabilization of humeralDecember 4, 2013
- FDASynthes 4.5 mm LCP Curved Condylar Plate Intended for buttressing multifragmentary distal femur fractures.December 4, 2013
Questions about this recall
Was Slipped Capital Femoral Epiphysis System Product Usage: Intended for fracture fixation of large bones and large bone fragments and for slipped capital femoral epiphysis, pediatric femoral neck fractures, intercondylar femur fractures and sacroiliac joint disruptions. recalled?
Yes. FDA published recall Z-0408-2014 on December 4, 2013. Reported hazard: There is a possibility for the Guide Wires of the Slipped Capital Femoral Epiphysis System to stick in the cannulation of the drill bit and the drill bits to break during surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter