MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Published January 31, 2018 · Recall initiated: December 21, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.
- Lot
- UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States