MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Published January 31, 2018 · Recall initiated: December 21, 2017
Summary
FDA records medical device Class II recall Z-0411-2018, published January 31, 2018 and initiated December 21, 2017. Recalling firm: Baxter Healthcare Corporation; Status: Terminated; Reported quantity: 186 units; Lot: UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.
- Lot
- UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDATruSystem 7000 U Operating Table (dV), Product Code 1723633November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDASPIRIVA HandiHaler (tiotropium bromide inhalation powder) 18 mcg/capsule (90 capsules (unit dose blisters) perFebruary 7, 2018
- FDAPURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000SJanuary 31, 2018
Questions about this recall
Was MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541 recalled?
Yes. FDA published recall Z-0411-2018 on January 31, 2018. Reported hazard: Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
What should I do if I have this product?
Voluntary: Firm initiated · Letter