Back to search
FDAFDA · Food and drugs·Recall number Z-0414-2020

Randox Liquid Cardiac Controls Catalogue Numbers CQ5053.

Published November 27, 2019 · Recall initiated: October 4, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Randox Laboratories, Limited
Distributor
US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
Lot
Catalogue Number: CQ5053 GTIN: 05055273207460 Batch/lot: 4245CK (Exp. 28Nov2019), 4248CK (Exp. 28Nov2019), 4313CK (Exp. 28Jun2020) and 4316CK (Exp. 28Jun2020)
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card