IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.

Published November 28, 2012 · Recall initiated: October 19, 2012

Summary

FDA records medical device Class II recall Z-0428-2013, published November 28, 2012 and initiated October 19, 2012. Recalling firm: Innovative Magnetic Resonance Imaging Systems Inc.; Status: Terminated; Reported quantity: 23; Lot: 10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Innovative Magnetic Resonance Imaging Systems Inc.
Distributor
Nationwide Distribution, including the states of FL, GA, IN, MD,MA, MI, MN, MO, TN, UT, and VA.
Lot
10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513
Risk classification
Class II
Area (FDA)
device
Country
Canada

Related recalls

Questions about this recall

Was IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems. recalled?

Yes. FDA published recall Z-0428-2013 on November 28, 2012. Reported hazard: IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.

What should I do if I have this product?

Voluntary: Firm initiated · Letter