IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.
Published November 28, 2012 · Recall initiated: October 19, 2012
Summary
FDA records medical device Class II recall Z-0428-2013, published November 28, 2012 and initiated October 19, 2012. Recalling firm: Innovative Magnetic Resonance Imaging Systems Inc.; Status: Terminated; Reported quantity: 23; Lot: 10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Innovative Magnetic Resonance Imaging Systems Inc.
- Distributor
- Nationwide Distribution, including the states of FL, GA, IN, MD,MA, MI, MN, MO, TN, UT, and VA.
- Lot
- 10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Canada
Related recalls
- FDAKING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 4 / Adult ClearSealNovember 28, 2012
- FDAMinotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High).November 28, 2012
- FDAKING LAD Flexible ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case "Size 3 / Child ClearSealNovember 28, 2012
- FDA0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride isNovember 28, 2012
- FDAKING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 20 devices/case Size 2 / Infant ClearSealNovember 28, 2012
- FDAStyker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single useNovember 28, 2012
Questions about this recall
Was IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems. recalled?
Yes. FDA published recall Z-0428-2013 on November 28, 2012. Reported hazard: IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.
What should I do if I have this product?
Voluntary: Firm initiated · Letter