Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge.
Published November 28, 2012 · Recall initiated: July 26, 2012
Summary
FDA records medical device Class II recall Z-0431-2013, published November 28, 2012 and initiated July 26, 2012. Recalling firm: Stryker Orthobiologics, Inc.; Status: Terminated; Reported quantity: 3,268; Lot: Part number 2113-0013, lot numbers T1203002, T1204005, and T1206001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step. Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe. T
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Orthobiologics, Inc.
- Distributor
- Nationwide Distribution-including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV
- Lot
- Part number 2113-0013, lot numbers T1203002, T1204005, and T1206001.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride isNovember 28, 2012
- FDA4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system.November 28, 2012
- FDAMinotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High).November 28, 2012
- FDASenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use withNovember 28, 2012
- FDAIMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the NeuroNovember 28, 2012
- FDASenTec V-Sign Membrane Changer Insert (VS-MCI), 5 pieces. The SenTec V-Sign Membrane Changer Insert isNovember 28, 2012
Questions about this recall
Was Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge. recalled?
Yes. FDA published recall Z-0431-2013 on November 28, 2012. Reported hazard: When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step. Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe. T
What should I do if I have this product?
Voluntary: Firm initiated · Letter