TB-0545PC: THUNDERBEAT 5 mm , 45 em, Pistol Grip TB-0535PC: THUNDERBEAT 5 mm, 35 em, Pi stol Grip TB-0545IC: THUNDERBEAT 5 mm, 45 em, In line Grip TB-0535IC: THUNDERBEAT 5 mm, 35 em, Inline Grip TB-0520IC: THUNDERBEAT 5 mm, 20 em, In line Grip TB-051 OIC: THUNDERBEAT 5 mm, I 0 em, In line Grip TB-545FC: THUNDERBEAT 5 mm, 45 em, Front Actuated Grip TB-0535FC: THUNDERBEAT 5 mm, 35 em, Front Actuated Grip The Olympus Thunderbeat Surgical Tissue Management System ("Thunderbeat") is intended to be used for open, laparoscopic (including single-site surgery), and endoscopic surgery to cut (dissect) or coagulate soft tissue or to ligate (seal and cut) vessels.
Published December 11, 2013 · Recall initiated: September 27, 2013
Summary
FDA records medical device Class II recall Z-0432-2014, published December 11, 2013 and initiated September 27, 2013. Recalling firm: Olympus America Inc.; Status: Terminated; Reported quantity: 8709; Lot: All sold units. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints of damage to various models of Thunderbeat Hand Instruments during surgical procedures, including cracking, breaking, or deformation of the components of the probe tip or jaw.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus America Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally to Canada.
- Lot
- All sold units
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEasy Diagnost Eleva Universal Fluoroscopic applicationsDecember 11, 2013
- FDAsyngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple labDecember 11, 2013
- FDADimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using softwareDecember 11, 2013
- FDASTERITITE Containers: STERITITE Container Kit, 1/4 size, 3" high, product code: 13902; STERITITE Mini MidDecember 11, 2013
- FDAOlympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and otherNovember 27, 2013
- FDAOlympus SurgMaster UES-40 electrosurgical unit ("UES-40") Product Usage: Designed for use in a medicalMay 15, 2013
Questions about this recall
Was TB-0545PC: THUNDERBEAT 5 mm , 45 em, Pistol Grip TB-0535PC: THUNDERBEAT 5 mm, 35 em, Pi stol Grip TB-0545IC: THUNDERBEAT 5 mm, 45 em, In line Grip TB-0535IC: THUNDERBEAT 5 mm, 35 em, Inline Grip TB-0520IC: THUNDERBEAT 5 mm, 20 em, In line Grip TB-051 OIC: THUNDERBEAT 5 mm, I 0 em, In line Grip TB-545FC: THUNDERBEAT 5 mm, 45 em, Front Actuated Grip TB-0535FC: THUNDERBEAT 5 mm, 35 em, Front Actuated Grip The Olympus Thunderbeat Surgical Tissue Management System ("Thunderbeat") is intended to be used for open, laparoscopic (including single-site surgery), and endoscopic surgery to cut (dissect) or coagulate soft tissue or to ligate (seal and cut) vessels. recalled?
Yes. FDA published recall Z-0432-2014 on December 11, 2013. Reported hazard: Complaints of damage to various models of Thunderbeat Hand Instruments during surgical procedures, including cracking, breaking, or deformation of the components of the probe tip or jaw.
What should I do if I have this product?
Voluntary: Firm initiated · Letter