Back to search
FDAFDA · Food and drugs·Recall number Z-0435-2013

100/120V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Published December 5, 2012 · Recall initiated: November 14, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be limi

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV.
Lot
Catalog number: 801763 and all serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card