100/120V AC, Advanced Perfusion System Platform (APS) The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Published December 5, 2012 · Recall initiated: November 14, 2012
Summary
FDA records medical device Class II recall Z-0435-2013, published December 5, 2012 and initiated November 14, 2012. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 1647 units total; Lot: Catalog number: 801763 and all serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be limi
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV.
- Lot
- Catalog number: 801763 and all serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPMJuly 2, 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974April 10, 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, SterileDecember 23, 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19REAugust 7, 2019
- FDAMedpro Elastomeric Infusion Pump. AccuFlux, Model # CT-0020-100H. Product Usage: The devices are intended forDecember 5, 2012
- FDA220/240V AC, Advanced Perfusion System Platform (APS) The Terumo® Advanced Perfusion System 1 is indicated forDecember 5, 2012
Questions about this recall
Was 100/120V AC, Advanced Perfusion System Platform (APS) The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. recalled?
Yes. FDA published recall Z-0435-2013 on December 5, 2012. Reported hazard: Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be limi
What should I do if I have this product?
Voluntary: Firm initiated · Letter