Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy

Published December 5, 2012 · Recall initiated: October 31, 2012

Summary

FDA records medical device Class II recall Z-0439-2013, published December 5, 2012 and initiated October 31, 2012. Recalling firm: Cooper Surgical, Inc.; Status: Terminated; Reported quantity: 1215 total; Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cooper Surgical, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
Lot
All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy recalled?

Yes. FDA published recall Z-0439-2013 on December 5, 2012. Reported hazard: Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle

What should I do if I have this product?

Voluntary: Firm initiated · Letter