FDAFDA · Food and drugs·Recall number Z-0450-2022
Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-M and H1-M-INT., Multi-vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The HawkOne catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The HawkOne catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.
Published January 26, 2022 · Recall initiated: December 6, 2021
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- ev3 Inc.
- Distributor
- Worldwide Distribution: US (nationwide), including PR; and Foreign distribution to countries of: Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Jordan, Korea, Republic Of Macao, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Arab Emirates, and United Kingdom.
- Lot
- H1-M - Irvine GTIN 00643169792302, Galway GTIN 00643169968332. H1-M-INT - Irvine GTIN 00643169728776, Galway, GTIN 00643169968523.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States