Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-M and H1-M-INT., Multi-vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The HawkOne catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The HawkOne catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.

Published January 26, 2022 · Recall initiated: December 6, 2021

Summary

FDA records medical device Class I recall Z-0450-2022, published January 26, 2022 and initiated December 6, 2021. Recalling firm: ev3 Inc.; Status: Terminated; Reported quantity: 71,260 units; Lot: H1-M - Irvine GTIN 00643169792302, Galway GTIN 00643169968332. H1-M-INT - Irvine GTIN 00643169728776, Galway, GTIN 00643169968523. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ev3 Inc.
Distributor
Worldwide Distribution: US (nationwide), including PR; and Foreign distribution to countries of: Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Jordan, Korea, Republic Of Macao, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Arab Emirates, and United Kingdom.
Lot
H1-M - Irvine GTIN 00643169792302, Galway GTIN 00643169968332. H1-M-INT - Irvine GTIN 00643169728776, Galway, GTIN 00643169968523.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-M and H1-M-INT., Multi-vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The HawkOne catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The HawkOne catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature. recalled?

Yes. FDA published recall Z-0450-2022 on January 26, 2022. Reported hazard: Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.

What should I do if I have this product?

Voluntary: Firm initiated · Letter