Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Published December 24, 2014 · Recall initiated: September 4, 2014
Summary
FDA records medical device Class II recall Z-0451-2015, published December 24, 2014 and initiated September 4, 2014. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 10 (4 US, 6 OUS).; Lot: Mfg Lot or Serial # System ID 00000642076BU4 303695IR1 00000642193BU7 0004040462 00000646910BU0 314768SM10 00000647792BU1 904UFHNSP 00000646492BU9 XV386041GE8 00000646048BU9 S4160099 00000644611BU6 H2009VAS44 00000646493BU7 H0008VAS06 00000645027BU4 XV106184CX5 00000646491BU1 4184240. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
- Lot
- Mfg Lot or Serial # System ID 00000642076BU4 303695IR1 00000642193BU7 0004040462 00000646910BU0 314768SM10 00000647792BU1 904UFHNSP 00000646492BU9 XV386041GE8 00000646048BU9 S4160099 00000644611BU6 H2009VAS44 00000646493BU7 H0008VAS06 00000645027BU4 XV106184CX5 00000646491BU1 4184240
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAGE Cabinet X-ray systemsDecember 24, 2014
- FDADiscovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic andDecember 24, 2014
Questions about this recall
Was Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. recalled?
Yes. FDA published recall Z-0451-2015 on December 24, 2014. Reported hazard: Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
What should I do if I have this product?
Voluntary: Firm initiated · Letter