Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Published December 24, 2014 · Recall initiated: September 4, 2014
Summary
FDA records medical device Class II recall Z-0452-2015, published December 24, 2014 and initiated September 4, 2014. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 1 - OUS; Lot: Mfg Lot or Serial # System ID 00000646778BU1 190020RX26. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
- Lot
- Mfg Lot or Serial # System ID 00000646778BU1 190020RX26
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAInnova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic andDecember 24, 2014
- FDAInnova 4100IQ Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography,December 24, 2014
Questions about this recall
Was Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. recalled?
Yes. FDA published recall Z-0452-2015 on December 24, 2014. Reported hazard: Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
What should I do if I have this product?
Voluntary: Firm initiated · Letter