QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Published January 12, 2022 · Recall initiated: November 8, 2021
Summary
FDA records medical device Class II recall Z-0455-2022, published January 12, 2022 and initiated November 8, 2021. Recalling firm: Qiagen Sciences LLC; Status: Terminated; Reported quantity: 376 kits; Lot: LOT 210209 UDI (01)14053228038846(17)211127(10)210209(24)691223 Serial Numbers: 412092551 412092559 412092552 412092560 412092553 412092561 412092554 412092563 412092555 412092564 412092556 412092565 412092557 412092566 412092558 412092567. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Qiagen Sciences LLC
- Distributor
- Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
- Lot
- LOT 210209 UDI (01)14053228038846(17)211127(10)210209(24)691223 Serial Numbers: 412092551 412092559 412092552 412092560 412092553 412092561 412092554 412092563 412092555 412092564 412092556 412092565 412092557 412092566 412092558 412092567
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAQIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acidsJuly 24, 2024
- FDAEZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecularApril 17, 2024
- FDAMetoprolol Tartrate Tablets, USP 25 mg, 1000 - count bottle, Rx Only, Distributed by: TruPharma, LLC. Tampa,January 19, 2022
- FDAMacaroni & Cheese, Macarrones con Queso; 3-5 lb pouches/bolsas, NET WT/PESO NETO 240 OZ (15 LB) 6.80 kg;January 12, 2022
- FDAtherascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121June 17, 2020
- FDAQIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument isAugust 28, 2019
Questions about this recall
Was QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223 recalled?
Yes. FDA published recall Z-0455-2022 on January 12, 2022. Reported hazard: The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
What should I do if I have this product?
Voluntary: Firm initiated · Letter