FDAFDA · Food and drugs·Recall number Z-0455-2022

QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223

Published January 12, 2022 · Recall initiated: November 8, 2021

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Qiagen Sciences LLC
Distributor
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
Lot
LOT 210209 UDI (01)14053228038846(17)211127(10)210209(24)691223 Serial Numbers: 412092551 412092559 412092552 412092560 412092553 412092561 412092554 412092563 412092555 412092564 412092556 412092565 412092557 412092566 412092558 412092567
Risk classification
Class II
Area (FDA)
device
Country
United States