Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.

Published December 3, 2014 · Recall initiated: October 23, 2014

Summary

FDA records medical device Class II recall Z-0463-2015, published December 3, 2014 and initiated October 23, 2014. Recalling firm: Abbott Molecular; Status: Terminated; Reported quantity: 9 CD ROMs (US); 1,273 CD ROMs (Foreign); Lot: List Number: 09K14; Models: E series, G series. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Molecular
Distributor
Worldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria , Australia , Austria , Belgium , Bosnia and Herzegovina , Cameroon , Canada , Chad , Colombia , Costa Rica , Croatia , Denmark , Egypt , Ethiopia , Finland , France , Germany , Greece , Guinea , Honduras , Hong Kong , India , Indonesia , Ireland , Israel , Italy , Japan , Kenya , Korea , Lithuania , Mali , Mexico , Netherlands , New Zealand , Nigeria , Norway , Pakistan , Paraguay , Poland , Portugal , Romania , Russia , Saudi Arabia , Senegal , Slovenia , South Africa , Spain , Sweden , Switzerland , Taiwan , Tanzania , Thailand , Turkey , Uganda , Ukraine , United Arab Emirates, United Kingdom , Vietnam
Lot
List Number: 09K14; Models: E series, G series
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing. recalled?

Yes. FDA published recall Z-0463-2015 on December 3, 2014. Reported hazard: Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.

What should I do if I have this product?

Voluntary: Firm initiated · Letter