Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.
Published December 3, 2014 · Recall initiated: October 23, 2014
Summary
FDA records medical device Class II recall Z-0463-2015, published December 3, 2014 and initiated October 23, 2014. Recalling firm: Abbott Molecular; Status: Terminated; Reported quantity: 9 CD ROMs (US); 1,273 CD ROMs (Foreign); Lot: List Number: 09K14; Models: E series, G series. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Molecular
- Distributor
- Worldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria , Australia , Austria , Belgium , Bosnia and Herzegovina , Cameroon , Canada , Chad , Colombia , Costa Rica , Croatia , Denmark , Egypt , Ethiopia , Finland , France , Germany , Greece , Guinea , Honduras , Hong Kong , India , Indonesia , Ireland , Israel , Italy , Japan , Kenya , Korea , Lithuania , Mali , Mexico , Netherlands , New Zealand , Nigeria , Norway , Pakistan , Paraguay , Poland , Portugal , Romania , Russia , Saudi Arabia , Senegal , Slovenia , South Africa , Spain , Sweden , Switzerland , Taiwan , Tanzania , Thailand , Turkey , Uganda , Ukraine , United Arab Emirates, United Kingdom , Vietnam
- Lot
- List Number: 09K14; Models: E series, G series
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrightView product code: 882478 BrightView X product code: 882480 BrightView XCT product code: 882482 andDecember 3, 2014
- FDAAlignRT- Intended for prescription use. The system is indicated for use during simulation, setup andDecember 3, 2014
- FDA5 mL Reaction Vessels (a consumable of the Abbott m2000sp Automated Fluid Handling System). The Abbott m2000August 20, 2014
- FDAThe CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR)April 30, 2014
- FDAAbbott m2000sp Instrument; an automated fluid handling system for performing sample preparation for nucleicJanuary 9, 2013
- FDAAbbott Molecular Bar Code Scanner User's Guide; instructions for setting up the Bar Code Scanner LN 6L89-01, aJuly 25, 2012
Questions about this recall
Was Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing. recalled?
Yes. FDA published recall Z-0463-2015 on December 3, 2014. Reported hazard: Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter