NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Published December 12, 2012 · Recall initiated: October 18, 2012
Summary
FDA records medical device Class II recall Z-0466-2013, published December 12, 2012 and initiated October 18, 2012. Recalling firm: Philips And Neusoft Medical Systems Co., Ltd.; Status: Terminated; Reported quantity: 20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series; Lot: NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips And Neusoft Medical Systems Co., Ltd.
- Distributor
- Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
- Lot
- NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- China
Related recalls
- FDANeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computedMay 1, 2013
- FDANeuViz Dual series CT Scanner System, Part number (PN): 989605651321. This is a computed X-ray, TomographyMay 1, 2013
- FDASiemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing SystemDecember 12, 2012
- FDABiOM Rechargeable Li Ion 22.2 V Battery and BiOM Lithium Battery Charger sold as an accessory to the BiOMDecember 12, 2012
- FDACustom Procedural Trays: Merit's Custom Total Hip Pack, K12T-05537, Merit's Total Knee Pack, K12T05538 ADecember 12, 2012
- FDANeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part numberAugust 1, 2012
Questions about this recall
Was NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. recalled?
Yes. FDA published recall Z-0466-2013 on December 12, 2012. Reported hazard: The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
What should I do if I have this product?
Voluntary: Firm initiated · Letter