FDAFDA · Food and drugs·Recall number Z-0466-2013
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Published December 12, 2012 · Recall initiated: October 18, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips And Neusoft Medical Systems Co., Ltd.
- Distributor
- Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
- Lot
- NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- China