Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis.
Published December 18, 2013 · Recall initiated: October 9, 2013
Summary
FDA records medical device Class II recall Z-0473-2014, published December 18, 2013 and initiated October 9, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 2912; Lot: J5606-C. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) Canada and Mexico
- Lot
- J5606-C
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDACoulter LH 780 Hematology Analyzer, Part No. 723585, A90728 Product Usage: Coulter LH 780 Hematology AnalyzerDecember 18, 2013
- FDACoulter LH 750 Hematology Analyzer, Part No. 6605632, A85570 Product Usage: The Coulter LH 750 HematologyDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDASynthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height ParallelDecember 11, 2013
Questions about this recall
Was Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis. recalled?
Yes. FDA published recall Z-0473-2014 on December 18, 2013. Reported hazard: A labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.
What should I do if I have this product?
Voluntary: Firm initiated · Letter