Philips Digital Diagnost Software Version R2.0.2 Product Usage: This system is used for making X-ray exposures for diagnostics
Published December 12, 2012 · Recall initiated: October 30, 2012
Summary
FDA records medical device Class II recall Z-0475-2013, published December 12, 2012 and initiated October 30, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 623 units; Lot: All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1 Models 712020, 712022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, India, Indonesia, Iraq, Italy, Japan, Latvia, Lithuania, Libya, Malta, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and UK.
- Lot
- All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1 Models 712020, 712022
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDABio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS KneeDecember 12, 2012
- FDASiemens Mobilette Mira, intended use as Mobile x-ray system.December 12, 2012
Questions about this recall
Was Philips Digital Diagnost Software Version R2.0.2 Product Usage: This system is used for making X-ray exposures for diagnostics recalled?
Yes. FDA published recall Z-0475-2013 on December 12, 2012. Reported hazard: With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
What should I do if I have this product?
Voluntary: Firm initiated · Letter