FDAFDA · Food and drugs·Recall number Z-0482-2021

ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Published December 2, 2020 · Recall initiated: October 20, 2020

Summary

FDA records medical device Class II recall Z-0482-2021, published December 2, 2020 and initiated October 20, 2020. Recalling firm: Viewray, Inc.; Status: Terminated; Reported quantity: 35 units; Lot: All Lots/Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Viewray, Inc.
Distributor
Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark.
Lot
All Lots/Serial Numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. recalled?

Yes. FDA published recall Z-0482-2021 on December 2, 2020. Reported hazard: There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

What should I do if I have this product?

Voluntary: Firm initiated · Letter