Stryker Orthopaedics Cancellous Bone Screw Howmedica Osteonics Corp. Stryker France Both the 6.4 mm diameter, 25mm length, Cancellous Bone Screws (Cat no 2030-6525-1) and the 5.5mm diameter, 40 mm length Cancellous Bone Screws (Cat no 2029-5540-1) are cleared for use in Total Hip Arthroplasty (THA), when implanting Stryker's Hemispherical Cluster Hole Acetabular Shell Systems and Total Knee Arthroplasty (TKA) when implanting Stryker's Series 7000 Standard Tibial Tray systems. The use of bone screws, during both THA and TKA are intended as supplemental fixation.

Published December 18, 2013 · Recall initiated: October 1, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stryker received a report from the field indicating that a 5.5 mm Cancellous Bone Screw 40 mm was identified in packaging associated with a 6.5 mm Cancellous Bone Screw 25 mm.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Howmedica Osteonics Corp.
Distributor
Nationwide Distribution
Lot
1) 6.5 mm Cancellous Bone Screw 25mm Catalog No: 2030-6525-1 Lot code MMLNLA 2) 5.5 Cancellous Bone Screw 40 mm Catalog No: 2029-5540-1 Lot Code MMLMJ1
Risk classification
Class II
Area (FDA)
device
Country
United States