FDAFDA · Food and drugs·Recall number Z-0486-2025
Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z
Published November 27, 2024 · Recall initiated: October 16, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.
- Lot
- Model Number C-OM-10000Z. UDI-DI Number: 00607567100008. Batch Numbers: 3000277221, 3000286549, 3000315868, 3000317517, 3000322538, 3000339447, 3000363028, 3000270713, 3000275435, 3000275771, 3000280868, 3000284390, 3000303149, 3000321642, 3000342528, 3000336070, 3000322179, 3000337396, 3000273182, 3000282181, 3000337880, 3000353535, 3000364172, 3000357109, 3000276109, 3000338107, 3000354445, 3000355050, 3000352899, 3000324785, 3000360894, 3000370720, 3000323627, 3000346330, 3000356740, 3000278665, 3000344012, 3000366729, 3000361801, 3000400928, 3000363775, 3000311956, 3000343553, 3000367559, 3000377818, 3000341430, 3000367345, 3000336377, 3000316208, 3000381449, 3000380956, 3000403756, 3000321414, 3000361499, 3000407809, 3000398030, 3000387678, 3000320077, 3000277072, 3000283754, 3000296552, 3000320660, 3000339096, 3000347694, 3000348390, 3000352250, 3000279303, 3000277309, 3000282794, 3000283482, 3000350599, 3000351774, 3000351996, 3000365741, 3000369223, 3000369738, 3000393258, 3000323129, 3000273547, 3000366133, 3000391146, 3000273843, 3000280827, 3000295912, 3000346410, 3000401855, 3000384033, 3000382967, 3000378703, 3000363755, 3000316146, 3000295949, 3000398861, 3000394893, 3000380542, 3000361797, 3000350598, 3000320073, 3000299042, 3000295186, 3000288696, 3000284723, 3000274876, 3000363227, 3000285664.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States