Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra Ultrasonic Surgical Aspirator system. One of the tubing lumen transmits irrigation fluid from the console to the handpiece of the system, and the other tubing lumen transmits aspirated fluid and tissue back to the system console. The aspirator is intended to be used to facilitate the removal of cellular and other unwanted soft tissue. The system provides selective tissue disintegration with simultaneous irrigation and aspiration. The system has been designed for use by surgeons in the areas of neurosurgery and gastroenterology including laparoscopic procedures. Its use in other procedures is regarded as experimental and is thus subjecto to appropriate local regulatory approval

Published December 10, 2014 · Recall initiated: September 29, 2014

Summary

FDA records medical device Class II recall Z-0491-2015, published December 10, 2014 and initiated September 29, 2014. Recalling firm: Integra LifeSciences Corp.; Status: Terminated; Reported quantity: 6,358 tubing kits; Lot: Lot Numbers: 130849;1132912;1132913;1132914;1133101;1133102;1133103;1133104;1133105;1133495;1134005;1134006;1134007;1134120;1134121;1134122;1134124;1134126;1134128;1134130;1134131;1134132;1134133;1134620;1134621;1134669;1134670;1134671;1134671;1134694;1134695;1140357;1141561;1141562;1141645;1141754;1141563;1141564;1141647;1142352;1142353;1142409; & 1142410. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Integra LifeSciences Corp.
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Australia, Bulgaria, Canada, Germany, Great Britain, Greece, Hungary, Indonesia, Israel, Italy, Japan, Mexico, Portugal, Spain, Sweden, Thailand, Turkey, and Wales.
Lot
Lot Numbers: 130849;1132912;1132913;1132914;1133101;1133102;1133103;1133104;1133105;1133495;1134005;1134006;1134007;1134120;1134121;1134122;1134124;1134126;1134128;1134130;1134131;1134132;1134133;1134620;1134621;1134669;1134670;1134671;1134671;1134694;1134695;1140357;1141561;1141562;1141645;1141754;1141563;1141564;1141647;1142352;1142353;1142409; & 1142410
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra Ultrasonic Surgical Aspirator system. One of the tubing lumen transmits irrigation fluid from the console to the handpiece of the system, and the other tubing lumen transmits aspirated fluid and tissue back to the system console. The aspirator is intended to be used to facilitate the removal of cellular and other unwanted soft tissue. The system provides selective tissue disintegration with simultaneous irrigation and aspiration. The system has been designed for use by surgeons in the areas of neurosurgery and gastroenterology including laparoscopic procedures. Its use in other procedures is regarded as experimental and is thus subjecto to appropriate local regulatory approval recalled?

Yes. FDA published recall Z-0491-2015 on December 10, 2014. Reported hazard: Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.

What should I do if I have this product?

Voluntary: Firm initiated · Letter