Back to search
FDAFDA · Food and drugs·Recall number Z-0496-2015

Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage

Published December 10, 2014 · Recall initiated: November 11, 2014

Summary

FDA records medical device Class II recall Z-0496-2015, published December 10, 2014 and initiated November 11, 2014. Recalling firm: Cure Medical LLC; Status: Terminated; Reported quantity: 192 cases = 1,920 boxes = 57,600 each; Lot: Lot Number: 140417-3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cure Medical LLC
Distributor
Nationwide Distribution
Lot
Lot Number: 140417-3
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage recalled?

Yes. FDA published recall Z-0496-2015 on December 10, 2014. Reported hazard: Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.

What should I do if I have this product?

Voluntary: Firm initiated · Letter