Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.
Published December 20, 2023 · Recall initiated: November 17, 2023
Summary
FDA records medical device Class I recall Z-0505-2024, published December 20, 2023 and initiated November 17, 2023. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 150 US; 192 ROW; Lot: All serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Domestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE NJ NV NY OH OK OR PA SC TN TX UT VA VT WA WI WV International distribution to Argentina Austria Canada China Colombia Croatia Czech Republic Denmark Ecuador Egypt France Germany Greece Italy Kuwait Libya Netherlands NO93 Panama Poland Portugal Qatar Russian Fed. Saint Kitts Saudi Arabia Spain Sweden Switzerland Syria Thailand T®rkiye United Kingdom Utd.Arab.Emir. Venezuela
- Lot
- All serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAGaw's Salami & Cheese Bagel 6oz, Foam Tray with Clear Plastic Wrap package, UPC:1725600501December 20, 2023
- FDAIconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor withDecember 20, 2023
Questions about this recall
Was Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device. recalled?
Yes. FDA published recall Z-0505-2024 on December 20, 2023. Reported hazard: Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter