Back to search
FDAFDA · Food and drugs·Recall number Z-0506-2025

Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.

Published November 27, 2024 · Recall initiated: October 18, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Zyno Medical LLC
Distributor
This product is distributed only in the United States.
Lot
Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.
Risk classification
Class II
Area (FDA)
device
Country
United States