AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Published February 14, 2018 · Recall initiated: August 24, 2017

Summary

FDA records medical device Class II recall Z-0508-2018, published February 14, 2018 and initiated August 24, 2017. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 265 units; Lot: 10094141 10848282 10848355. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution
Lot
10094141 10848282 10848355
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. recalled?

Yes. FDA published recall Z-0508-2018 on February 14, 2018. Reported hazard: Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.

What should I do if I have this product?

Voluntary: Firm initiated · Letter