Siemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Intended Use For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems.

Published December 19, 2012 · Recall initiated: November 14, 2012

Summary

FDA records medical device Class II recall Z-0511-2013, published December 19, 2012 and initiated November 14, 2012. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 100 Test=27392; 500 test=21045 and Reference Kits=3773; Lot: Al lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Operating range of the ADVIA Centaur systems is 18C® to 30®C (64.4®F to 86.0®F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Lot
Al lot codes
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens Healthcare ADVIA Centaur Folate (FOL) 500 Test 06891541 100 Test 06367974 Ref Kit (US only) 09132781 Ref Kit (US only) 00203473 Intended Use For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems. recalled?

Yes. FDA published recall Z-0511-2013 on December 19, 2012. Reported hazard: Operating range of the ADVIA Centaur systems is 18C® to 30®C (64.4®F to 86.0®F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

What should I do if I have this product?

Voluntary: Firm initiated · Letter