Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
Published December 25, 2013 · Recall initiated: October 2, 2013
Summary
FDA records medical device Class II recall Z-0519-2014, published December 25, 2013 and initiated October 2, 2013. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 16,154 units; Lot: Part Number 81901, Lot Numbers: 13181168-13341170. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, India, Hong Kong, Guatemala, China, Brazil, Brunei Darussalam, Costa Rica, El Salvador, JordanMalaysia, Mexico, Panama, Philippines, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Venezuela, and Viet Nam.
- Lot
- Part Number 81901, Lot Numbers: 13181168-13341170.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAKimberly-Clark Single Shot Epidural Pain Management Trays containing the BD LOR Syringe (BD Code 405198)December 25, 2013
- FDAMega Needle Driver, 8MM assembly of the da Vinci S IS1200 and da Vinci Si, IS2000. Manufactured by IntuitiveDecember 25, 2013
Questions about this recall
Was Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. recalled?
Yes. FDA published recall Z-0519-2014 on December 25, 2013. Reported hazard: Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I
What should I do if I have this product?
Voluntary: Firm initiated · Letter