FDAFDA · Food and drugs·Recall number Z-0522-2013
Cobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitro measurement of pH, blood gases (BG), electrolytes Na+ K+, iCa2+ (ISE), hematocrit (Hct), metabolites (Gluc, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHB, COHB, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory.
Published December 19, 2012 · Recall initiated: November 14, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a change in the thickness in the cuvette layer. Internal investigations found the altered thickness of the cuvette was due to production equipment settings. Cuvettes were produced within specifications, but within the upper
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide Distribution including NE and AZ.
- Lot
- Part number: 05169992001 cobas b 123 Fluid Pack COOX, 200 with Lot# 21426121 and Lot# 21426171.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States