Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.
Published December 25, 2013 · Recall initiated: November 13, 2013
Summary
FDA records medical device Class II recall Z-0527-2014, published December 25, 2013 and initiated November 13, 2013. Recalling firm: Karl Storz Endoscopy America Inc; Status: Terminated; Reported quantity: 141 units; Lot: C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Karl Storz Endoscopy America Inc
- Distributor
- Nationwide Distribution.
- Lot
- C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1,December 25, 2013
- FDAOxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mcg, 50 discs per cartridge, 5December 25, 2013
- FDAVITROS CA 125 II Calibrators For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems,December 25, 2013
- FDAAxillent Debakey Grasping Forceps, Part No. 33821DYD In general the surgeons use 33821DYD Debakey forceps toDecember 18, 2013
- FDA27023WU Balloon Catheter, 3 Fr., 2/pkg, sterile, contains latex. The RUSCH ureter occlusion catheter isJuly 24, 2013
- FDAElectrode, Roller Ball, 24FR (27050NK/6). Resectoscope accessories (ie. electrode, cutting loops and coldMay 22, 2013
Questions about this recall
Was Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants. recalled?
Yes. FDA published recall Z-0527-2014 on December 25, 2013. Reported hazard: KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
What should I do if I have this product?
Voluntary: Firm initiated · Letter