Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter)
Published February 9, 2022 · Recall initiated: December 2, 2021
Summary
FDA records medical device Class II recall Z-0532-2022, published February 9, 2022 and initiated December 2, 2021. Recalling firm: LivaNova USA, Inc.; Status: Terminated; Reported quantity: 10; Lot: UDI: 05425025750115. Serial Numbers: 119441, 119442, 119443, 119452, 119453, 119454, 119455, 119456, 119457, 119458. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LivaNova USA, Inc.
- Distributor
- US: NE, MO, AR, WV, KS, MA, IN
- Lot
- UDI: 05425025750115. Serial Numbers: 119441, 119442, 119443, 119452, 119453, 119454, 119455, 119456, 119457, 119458
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for VagusFebruary 12, 2025
- FDAWayfare Dairy Free Butter Salted & Whipped,February 9, 2022
- FDAThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrodeFebruary 9, 2022
- FDAtaste of Inspirations CREME PUFFS Whipped Cream Filled Pastry NET WT 13.2 OZ (375g) 30ct UPC: 7 25439 99597 9,February 9, 2022
- FDAThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrodeFebruary 9, 2022
- FDADelizza BELGIAN MINI CREAM PUFFS WHIPPED FILLING MADE WITH FRESH CREAM NET WT 13.2 oz (375g) PRODUCED IN THEFebruary 9, 2022
Questions about this recall
Was Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter) recalled?
Yes. FDA published recall Z-0532-2022 on February 9, 2022. Reported hazard: A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit