Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure
Published December 26, 2012 · Recall initiated: November 15, 2012
Summary
FDA records medical device Class II recall Z-0533-2013, published December 26, 2012 and initiated November 15, 2012. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 200; Lot: Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide distribution - USA (nationwide) and the countries of Australia, Canada, Chile, Europe, Japan, Korea, Panama, Puerto Rico and Thailand.
- Lot
- Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAMedline Total Knee CDS; a single patient prescription procedure pack and disposables, including a back tableDecember 26, 2012
- FDAArobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components ofDecember 26, 2012
Questions about this recall
Was Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure recalled?
Yes. FDA published recall Z-0533-2013 on December 26, 2012. Reported hazard: Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.
What should I do if I have this product?
Voluntary: Firm initiated · Letter