Medfusion® 4000 Syringe Infusion Pumps Medfusion® 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.

Published December 25, 2013 · Recall initiated: November 22, 2013

Summary

FDA records medical device Class II recall Z-0540-2014, published December 25, 2013 and initiated November 22, 2013. Recalling firm: Smiths Medical ASD, Inc.; Status: Terminated; Reported quantity: 9560 (8601 US, 959 OUS); Lot: All Medfusion® 4000 Syringe Pumps in the field are affected by this Field Safety Corrective Action. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Smiths Medical is conducting a recall involving all Medfusion® 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD, Inc.
Distributor
Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
Lot
All Medfusion® 4000 Syringe Pumps in the field are affected by this Field Safety Corrective Action.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medfusion® 4000 Syringe Infusion Pumps Medfusion® 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes. recalled?

Yes. FDA published recall Z-0540-2014 on December 25, 2013. Reported hazard: Smiths Medical is conducting a recall involving all Medfusion® 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po

What should I do if I have this product?

Voluntary: Firm initiated · Letter