Back to search
FDAFDA · Food and drugs·Recall number Z-0545-2013

Persona, The Personalized Knee System,Tibial Drill, Cemented, 15.7 mm Diameter, Nonsterile This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.

Published December 26, 2012 · Recall initiated: November 14, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
Lot
Part Number 42-5399-018-00,Lot 77001868
Risk classification
Class II
Area (FDA)
device
Country
United States