Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.
Published January 1, 2014 · Recall initiated: December 10, 2013
Summary
FDA records medical device Class II recall Z-0570-2014, published January 1, 2014 and initiated December 10, 2013. Recalling firm: Philips Medical Systems, Inc.; Status: Terminated; Reported quantity: 18,529 units; Lot: Software versions: F.02, R.02, 9.xx or any earlier versions Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems, Inc.
- Distributor
- US Nationwide and OUS.
- Lot
- Software versions: F.02, R.02, 9.xx or any earlier versions Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFoot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 ThisApril 6, 2016
- FDAPhilips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAIMMULITE/IMMULITE 1000 Systems FOL Folic Acid, REF/Catalog Number LKFO1 (100 tests), Siemens Material NumberJanuary 1, 2014
- FDAInstrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versionsJanuary 1, 2014
Questions about this recall
Was Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display. recalled?
Yes. FDA published recall Z-0570-2014 on January 1, 2014. Reported hazard: MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
What should I do if I have this product?
Voluntary: Firm initiated · Letter