Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 knee implant.
Published January 13, 2016 · Recall initiated: November 9, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
One lot of 35 mm patella implants is labeled as 32 mm. These implants have pegs which are spaced proportionally; therefore the mismatch prevents the pegs from properly aligning with pre-drilled holes during implantation.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Smith & Nephew, Inc.
- Distributor
- US Distribution to states of:FL, MO, IN and TX.
- Lot
- Batch # 15GM07831
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States